KMID : 0360220170580010021
|
|
Journal of the Korean Ophthalmological Society 2017 Volume.58 No. 1 p.21 ~ p.26
|
|
Comparison of Refractive Power and Astigmatism between Digital Keratometer and Autorefractor
|
|
Choi Jae-Won
Han Sang-Youp Lee Kyung-Heon
|
|
Abstract
|
|
|
Purpose: To compare the corneal refraction, astigmatism, and corneal marking for toric intraocular lens measured with a VERION¨Þ Image Guided System and an autorefractor in order to precisely determine the toric intraocular lens target.
Methods: The medical records of 29 eyes of 29 patients were retrospectively reviewed to compare the corneal refraction, axis, and amount of astigmatism measured with a VERION¨Þ Image Guided System and an autorefractor. We compared the difference in marked axis for toric intraocular lenses as measured by a VERION Digital Marker and a manual marker.
Results: The average corneal refraction and amount of astigmatism were greater with the VERION¨Þ Image Guided System than with the autorefractor. The average difference in axis of astigmatism was 9.62¡Æ. The difference in axis of astigmatism in patients with more than 1.0 D of astigmatism (18 eyes) was 2.82 ¡¾ 2.27¡Æ, while that in patients with less than 1.0 D of astigmatism (11 eyes) was 20.27 ¡¾ 28.14¡Æ. The average difference in marked axis for toric intraocular lens measured by the VERION¨Þ Digital Marker and manual marker was 2.50¡Æ (0¡Æ-9.27¡Æ).
Conclusions: The corneal refraction and amount of astigmatism were significantly higher with the VERION¨Þ Image Guided System, so careful concern and comparison of surgical outcomes between the two devices is needed in biometry.
|
|
KEYWORD
|
|
Astigmatism, Autorefracto, Corneal refraction, Digital marker, VERION¨Þ
|
|