Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0360220170580030276
Journal of the Korean Ophthalmological Society
2017 Volume.58 No. 3 p.276 ~ p.282
Comparison of Central Corneal Thickness Measured by Swept-source Optical Coherence Tomography and Ultrasound Pachymetry
Shin Woo-Beom

Jeong Hyun-Kyo
Kim Min-Kyo
Kim Hong-Seok
Lee Jun-Mo
Kim Chan-Yun
Kim Ji-Hyun
Abstract
Purpose: To analyze and compare ultrasound pachymetry (USP) with a more recently adopted device, the intra ocular lens (IOL) master 700, which are both used to measure central corneal thickness.

Methods: The central corneal thickness was measured in 24 eyes of 12 glaucoma patients and in 83 eyes of 42 normal patients. First, the IOL master 700 was used to measure the central corneal thickness, followed by measurements taken using USP later. The results were analyzed using a paired t-test. We analyzed the agreement and the correlations between the two test devices by using Bland-Altman plots and the Pearson correlation test. To evaluate the reproducibility, measurements with the IOL master 700 were taken twice for a few normal patients within a small time interval.

Results: Via the IOL master 700, the thickness of the central cornea showed a high reproducibility and repeatability, demonstrating 2.7 ¡¾ 1.7 ¥ìm for the test-retest variability, 6.78% for the coefficient of variation, and 0.997 for the intraclass correlation value. The mean measurements using USP and the IOL master 700 are 554.4 ¡¾ 37.4 ¥ìm and 551.1 ¡¾ 37.1 ¥ìm, respectively, showing that the IOL master 700 measured significantly smaller values than USP with a p-value < 0.001. The deviations between the two methods are scattered throughout the 95% confidence interval. According to the Pearson correlation test, the measured values ??of the two test devices were found to have a highly positive correlation (r = 0.977, p < 0.0001).

Conclusions: This study demonstrated that the central corneal thickness (CCT) measured via the IOL master was significantly thinner than that of USP, and the two test devices had a high correlation and good agreement. The CCT value measured via the IOL master 700 also exhibited high reproducibility.
KEYWORD
Central corneal thickness, Ultrasound pachymetry, IOL master 700
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø