Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0360220170580050523
Journal of the Korean Ophthalmological Society
2017 Volume.58 No. 5 p.523 ~ p.529
Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 700, IOL Master 500, and Ultrasound
Yoo Tae-Keun

Choi Moon-Jung
Lee Hyung-Keun
Seo Kyoung-Yul
Kim Eung-Kweon
Kim Tae-Im
Abstract
Purpose: To compare the new swept-source optical coherence tomography based IOL Master 700 to both the partial coherence interferometry based IOL Master 500 and ultrasound A-scan in terms of the ocular biometry and the prediction of postoperative refractive outcomes.

Methods: A total 67 eyes of 55 patients who received cataract surgery were included in our study. The axial length, anterior chamber depth, and keratometry were measured using IOL Master 700, IOL Master 500, and A-scan. The predictive errors, which are the differences between predictive refraction and post-operative refraction 1 month after surgery, were also compared.

Results: Axial length measurements were not successful in 5 eyes measured using IOL Master 700 and in 12 eyes measured using IOL Master 500. The mean absolute postoperative refraction predictive errors were 0.63 ¡¾ 0.50 diopters, 0.66 ¡¾ 0.51 diopters, and 0.62 ¡¾ 0.51 diopters for IOL Master 700, IOL Master 500, and A-scan, respectively, and these values exhibited no statistically significant differences. The mean axial lengths were 24.25 ¡¾ 2.41 mm, 24.24 ¡¾ 2.40 mm, and 24.22 ¡¾ 2.39 mm; the mean anterior chamber depths were 3.09 ¡¾ 0.39 mm, 3.17 ¡¾ 0.39 mm, and 3.15 ¡¾ 0.46 mm; and the mean keratometry values were 44.12 ¡¾ 1.82 diopters, 44.57 ¡¾ 2.10 diopters, and 43.98 ¡¾ 1.84 diopters for the IOL Master 700, IOL Master 500, and A-scan groups, respectively. None of these parameters showed statistically significant differences between the three groups. Regarding pair-wise comparison, there were significant differences between the IOL Master 700 and the other devices.

Conclusions: The ocular biometric measurements measured using IOL Master 700, IOL Master 500, and A-scan showed no significant differences. However, IOL Master 700 demonstrated a superior ability to successfully take biometric measurements compared to IOL Master 500. Therefore, IOL Master 700 is capable of measuring ocular biometry for cataract surgery in clinical practice.
KEYWORD
A-scan, Biometry, Cataract, Intraocular lens, IOL Master 700
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø