Comparison of Intraocular Lens Power Calculation Methods after Refractive Surgery Provided by the ASCRS
Jung So-Yeon

Jung Ji-Won
Abstract
Purpose: Methods of intraocular lens (IOL) calculations for cataract surgery after myopic corneal refractive surgery were compared.

Methods: The medical records of 23 eyes from 20 patients were retrospectively analyzed. Five methods - Haigis L and Barrett no history method from Post-Refractive IOL Calculator of American Society of Cataract and Refractive Surgery (ASCRS), Shammas no history method, Haigis suite from IOL Master 700, and clinical history method - were used.

Results: The Barrette no history method from the ASCRS showed the lowest arithmetic errors, while the Haigis-L from the ASCRS showed the lowest absolute prediction errors (-0.38 ¡¾ 0.99 diopters [D] and 0.76 ¡¾ 0.68 D, respectively). The percentages of refractive prediction errors within 0.50 D and 1.00 D of the Barrett no history method were 52.2% and 82.6%, respectively.

Conclusions: The Barrette no history method and Haigis-L from the ASCRS showed the smallest errors in this study. The Barrett no history method showed a high percentage of refractive prediction errors within 0.50 D and 1.00 D.
KEYWORD
American Society of Cataract and Refractive Surgery (ASCRS) intraocular lens (IOL) calculator, Cataract, Corneal refractive surgery, IOL power calculation