Journal of the Korean Ophthalmological Society
2022 Volume.63 No. 1 p.10 ~ p.19
Comparison of Anterior Segment Measurements between Scheimpflug- Placido Camera and New Swept-source Optical Coherence Tomography
Choe Gon-Soo

Kim Kook-Young
Abstract
Purpose: We compared anterior segment measurements obtained using a Scheimpflug-Placido topographer (SIRIUS¢ç, Costruzione Strumenti Oftalmici, Florence, Italy) and a new anterior module for a swept-source optical coherence tomography system (ANTERION¢ç, Heidelberg Engineering Inc., Heidelberg, Germany).

Methods: Anterior segment measurements were evaluated in 74 eyes of 101 patients with the two devices. Central corneal thickness (CCT), anterior chamber depth (ACD), corneal refractive power (K), J0 (Jackson cross-cylinder with axes at 180¡Æ and 90¡Æ), J45 vector (Jackson cross-cylinder with axes at 45¡Æ and 135¡Æ), and white-to-white corneal diameter (WTW) measurements obtained using the SIRIUS¢ç system and the new anterior segment module of ANTERION¢ç were compared.

Results: The mean CCTs measured by SIRIUS¢ç and ANTERION¢ç were 558.69 ¡¾ 44.65 and 540.26 ¡¾ 36.57 ¥ìm, respectively.
The difference was statistically significant (p < 0.005), and there were high correlations between the two methods (r = 0.915, p < 0.001). The mean ACD measurements were 3.34 ¡¾ 0.53 and 3.34 ¡¾ 0.51 mm (p = 0.856), respectively, for SIRIUS¢ç and ANTERION¢ç. The measurements were strongly correlated (r = 0.966, p < 0.001). The measurements of WTW, total mean K, and J0 obtained with the two systems differed significantly (p < 0.005, p = 0.017, and p < 0.005, respectively), with high correlations between the two systems (r = 0.875, r = 0.967, and r = 0.933, respectively; all p < 0.001).

Conclusions: There were significant differences in measurements of K, CCT, WTW, and J0 between the two devices.
Agreement analysis suggests that SIRIUS¢ç and ANTERION¢ç should not be used interchangeably.
KEYWORD
Anterior chamber depth, Central corneal thickness, Corneal refractive power, Swept-source optical coherence tomography