KMID : 0360220220630050426
Journal of the Korean Ophthalmological Society
2022 Volume.63 No. 5 p.426 ~ p.433
Accuracy of Intraocular Pressure Measurements of Eyes with Therapeutic Contact Lenses after Vitrectomy
Moon Ja-Young

Kim Yung-Hui
Ji Yong-Sok
Abstract
Purpose: To evaluate the accuracy of intraocular pressure (IOP) measurements obtained by a rebound and non-contact tonometer in eyes with a therapeutic contact lens (CL) after vitrectomy.

Methods: In 60 eyes of 60 patients who underwent vitrectomy for vitreoretinal disease, IOP was measured using a rebound tonometer (iCare ic200¢ç; IOPRT) and non-contact computerized air puff tonometer (CT-80, IOPNCT), before and after wearing a CL (Purevision2¢ç, +0.0 diopter). The mean IOP of three consecutive measurements were analyzed, and a comparative analysis with IOP measured by a Goldman applanation tonometer (IOPGAT) was performed.

Results: The mean IOPRT without and with the CL was 12.55 ¡¾ 5.43 and 13.12 ¡¾ 5.13 mmHg, respectively, showing a statistically significant difference (p = 0.02) and strong positive correlation (r = 0.90, p < 0.001). The mean IOPNCT with and without the CL was 12.18 ¡¾ 3.24 and 12.17 ¡¾ 3.14 mmHg, showing no statistically significant difference (p = 0.17). The consistency with IOPGAT (12.57 ¡¾ 5.22 mmHg) was highest in IOPRT without the CL, followed by IOPRT with the CL, IOPNCT without the CL, and IOPNCT with the CL (intraclass correlation coefficients = 0.955, 0.945, 0.856, and 0.850, respectively). In addition, the rebound tonometer successfully measured IOP, regardless of whether the CL was worn; however, the non-contact tonometer failed to measure IOP in seven eyes without the CL and nine with the CL. No difference was observed according to intraocular tamponade type.

Conclusions: A rebound tonometer can be used as an alternative IOL measuring method in eyes for which it is difficult to use a Goldman applanation tonometer due to the postoperative presence of a therapeutic CL.
KEYWORD
Intraocular pressure, Non-contact pneumotonometry, Rebound tonometry, Therapeutic contact lens, Vitrectomy