Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0360220240650020108
Journal of the Korean Ophthalmological Society
2024 Volume.65 No. 2 p.108 ~ p.116
Development of an Instrument for Slit-lamp Examination of Donor Corneas in Preservation Medium
Nam Ga-Hee

Kim Da-Ran
Yoon Young-Chae
Yang Soon-Won
Whang Woong-Joo
Byun Yong-Soo
Hwang Hyung-Bin
Na Kyung-Sun
Lee Hyun-Soo
Chung So-Hyang
Kim Eun-Chul
Cho Yang-Kyung
Kim Hyun-Seung
Hwang Ho-Sik
Abstract
Purpose: To evaluate the effectiveness of an instrument devised for slit-lamp examination of donor corneas suspended in preservation medium.

Methods: The study examined two donor corneas received at Yeouido St. Mary's Hospital in February 2023 and March 2023.
The instrument has three main components: a plastic holder to hold the preservation medium bottle, a cube with a mirror for reflecting the slit beam, and a stand to attach the device to the slit-lamp. Using the instrument, the donor corneas were examined via slit-lamp: microscopy with the endothelium facing upward and downward. Specular microscopy and anterior segment optical coherence tomography (OCT) were also performed on the preserved donor corneas.

Results: Slit-lamp examination of donor corneas in preservation medium using the instrument showed overall corneal buttoning and optical sections of the donor cornea. Using specular reflection and retroillumination, the endothelial layer was partially visible. However, specular microscopy and anterior segment OCT could not examine the donor cornea in preservation medium using the instrument.

Conclusions: The devised instrument facilitates slit-lamp examination of donor corneas in preservation medium, enabling a qualitative assessment of donor corneas before corneal transplantation surgery.
KEYWORD
Corneal transplantation, Donor cornea, Slit-lamp examination
FullTexts / Linksout information
 
Listed journal information