KMID : 0361620060410040687
|
|
Journal of the Korean Orthopaedic Association 2006 Volume.41 No. 4 p.687 ~ p.694
|
|
The Harris-Galante-II Acetabular Componentin Cementless Total Hip Arthroplasty- Average Ten-year Follow-up Results -
|
|
Chang Jun-Dong
Yoo Je-Hyun Lee Sang-Soo Kwon Soon-Eok
|
|
Abstract
|
|
|
Purpose: To evaluate the average ten-year follow-up results of 98 total hip arthroplasties using Harris-Galante-II porous-coated acetabular components.
Materials and Methods: Ninety-eight hips were evaluated. The mean age of the patients was 45 years, and the mean follow-up was 121 months. The clinical and radiographic findings including the wear of the polyethylene liner were evaluated. The endurance of the acetabular component was analyzed using the Kaplan-Meier method.
Results: Retroacetabular osteolysis, dissociation of the PE liner and aseptic loosening of the acetabular component was observed in 25 hips (25.5%), 9 hips (9.2%), and 1 hip (1.0%), respectively. The average rate of liner wear was 0.16 mm/year. Twenty acetabular components (20.4%) were revised, and the survival probability of the acetabular components at 10 years was 87% (95% confidence limits, 83-90%).
Conclusion: In this long-term follow-up study, the revision rate of Harris-Galante-II acetabular component was relatively high, and was associated with locking mechanism failure and retroacetabular osteolysis. However, the rarity of aseptic loosening suggests that excellent long-term fixation could be achieved with the surface treatment of a fiber-metal porous coating.
|
|
KEYWORD
|
|
Total hip arthroplasty, Harris-Galante-II acetabular component
|
|