KMID : 0361620060410040695
|
|
Journal of the Korean Orthopaedic Association 2006 Volume.41 No. 4 p.695 ~ p.702
|
|
Minimum 4 Year Follow-up Results of Metalon Metal Hip Resurfacing Arthroplasty
|
|
Lee Soo-Ho
Cho Soo-Hyun Kang Seok-Joong Nam Tae-Seok Choi Ji-Won
|
|
Abstract
|
|
|
Purpose: To report the minimum 4 year follow up results of metal on metal hip resurfacing arthroplasty.
Materials and Methods: 39 hips from 36 patients, who received hip resurfacing from November 1998 to September 2001 and were followed up for at least 4 years, were enrolled in this study. The average age was 45 years (28-69 years) and the average follow up period was 51.9 months (48-82 months). A clinical evaluation was performed with the Harris Hip Score and gait analysis postoperatively. A radiologic evaluation was performed regularly after surgery using anteroposterior and lateral simple radiographs. The post-operative cobalt-chrome serum concentration was measured and compared using 21 cases of ceramic-on-ceramic total hip arthroplasty as a control group.
Results: The Harris Hip Score improved from 61 points (31-74) preoperatively to 96 points (85-99) postoperatively. There were no cases of acetabular loosening, and the average angle of femoral insert to the shaft was 136 degrees with 22 hips (61.1%) being within 130-140 degrees. The 4 year follow-up survival rate was 97.4%. The serum cobalt and chrome levels were higher than the control group (p<0.05).
Conclusion: Hip resurfacing showed excellent survival rate after a short-term follow-up, but still requires a long-term evaluation with more cases and further metal ion release studies.
|
|
KEYWORD
|
|
Hip resurfacing arthroplasty, Metal ion, Cobalt, Chrome
|
|