Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0361620140490020172
Journal of the Korean Orthopaedic Association
2014 Volume.49 No. 2 p.172 ~ p.176
Mechanical Failure in Distraction Osteogenesis Using the Intramedullary Skeletal Kinetic Distractor
Lee Yun-Seok

Kim Bang-Hyun
Lee Dong-Hoon
Abstract
In an effort to overcome the drawbacks of distraction osteogenesis using the Ilizarov external fixator, intramedullary lengthening devices have been developed and applied for long bone distraction osteogenesis. Several successful cases have been reported, leading to the next generation of distraction osteogenesis. However, intramedullary lengthening devices have their own problems, such as device failure, difficulty of control of the lengthening degree. The authors report on a case of device failure during distraction osteogenesis using ISKD¢ç (Orthofix Inc.) in the distraction phase, and the strategy that involved switching to a lengthening over nail system, which uses the Ilizarov external fixator, in order to obtain the target length.
KEYWORD
equipment failure analysis, leg length inequality/surgery, distraction osteogenesis/instrumentation