Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0361820080270040229
Journal of the Korean Society of Medical Ultrasound
2008 Volume.27 No. 4 p.229 ~ p.232
A New Ultrasound-Guided Puncture Device with Augmented Degrees of Performance Freedom and Ability to Attach to Most Convex Probes
Jung Sung-Il

Son Kyu-Ri
Kim Young-Jun
Jeon Hae-Jeong
Park Sang-Woo
Cho Jeong-Yeon
Kim Seung-Hyup
Abstract
Purpose: To evaluate the performance of a newly-designed ultrasound (US)-guided puncture device.

Materials and Method : A newly-designed US-guided puncture device was composed of a guide segment and an attachable segment. The guide segment allowed the needle to be placed in the plane of US view with a maximal degree of freedom, and the attachable segment was designed to attach to most convex US probes. Six operators punctured 144 targets in phantoms using either the new device (n = 72) or free-hand technique (n = 72). The number of required needle passages and the necessary procedure times were compared between the two groups.

Results: The number of required needle passages and the necessary procedure time were significantly reduced in five operators when the newly-designed US-guided puncture device was used (p < 0.05).

Conclusion: A newly-designed US-guided puncture device, which allows for a maximal degree of freedom in needle placement and can attach to most convex US probes, showed good performance in our study.
KEYWORD
Ultrasound, Interventional procedure, Guidance, Technology
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø