KMID : 0361820080270040229
|
|
Journal of the Korean Society of Medical Ultrasound 2008 Volume.27 No. 4 p.229 ~ p.232
|
|
A New Ultrasound-Guided Puncture Device with Augmented Degrees of Performance Freedom and Ability to Attach to Most Convex Probes
|
|
Jung Sung-Il
Son Kyu-Ri Kim Young-Jun Jeon Hae-Jeong Park Sang-Woo Cho Jeong-Yeon Kim Seung-Hyup
|
|
Abstract
|
|
|
Purpose: To evaluate the performance of a newly-designed ultrasound (US)-guided puncture device.
Materials and Method : A newly-designed US-guided puncture device was composed of a guide segment and an attachable segment. The guide segment allowed the needle to be placed in the plane of US view with a maximal degree of freedom, and the attachable segment was designed to attach to most convex US probes. Six operators punctured 144 targets in phantoms using either the new device (n = 72) or free-hand technique (n = 72). The number of required needle passages and the necessary procedure times were compared between the two groups.
Results: The number of required needle passages and the necessary procedure time were significantly reduced in five operators when the newly-designed US-guided puncture device was used (p < 0.05).
Conclusion: A newly-designed US-guided puncture device, which allows for a maximal degree of freedom in needle placement and can attach to most convex US probes, showed good performance in our study.
|
|
KEYWORD
|
|
Ultrasound, Interventional procedure, Guidance, Technology
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|