Korean Journal of Orthodontics
2008 Volume.38 No. 4 p.228 ~ p.239
Cortical bone strain during the placement of orthodontic microimplant studied by 3D finite element analysis
Nam Ok-hyun

Yu Won-Jae
Kyung Hee-Moon
Abstract
Objective: The aim of this study was to evaluate the strain induced in the cortical bone surrounding an orthodontic microimplant during insertion.

Methods: A 3D finite element method was used to model the insertion of a microimplant (AbsoAnchor SH1312-7, Dentos Co., Daegu, Korea) into 1 mm thick cortical bone with a pre-drilled hole of 0.9 mm in diameter. A total of 1,800 analysis steps was used to simulate the 10 turns and 5 mm advancement of the microimplant. A series of remesh in the cortical bone was allowed to accommodate the change in the geometry accompanied by the implant insertion.

Results: Bone strains of well higher than 4,000 microstrain, the reported upper limit for normal bone remodeling, was observed in the bone along the whole length of the microimplant. At the bone in the vicinity of the screw tip, strains of higher than 100% was recorded. The insertion torque was calculated at approximately 1.2 Ncm which was slightly lower than those measured from the animal experiment using rabbit tibias.

Conclusions: The insertion process of a microimplant was successfully simulated using the 3D finite element method which showed that bone strains from a microimplant insertion might have a negative impact on physiological remodeling of bone. (Korean J Orthod 2008;38(4):228-239)
KEYWORD
Microimplant, Strainduringinsertion, 3Dfiniteelementmethod, Rabbitexperiment