Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0361920170470040238
Korean Journal of Orthodontics
2017 Volume.47 No. 4 p.238 ~ p.247
Effects of recycling on the biomechanical characteristics of retrieved orthodontic miniscrews
Yun Soon-Dong

Choi Sung-Hwan
Cha Jung-Yul
Yu Hyung-Seog
Kim Kwang-Mahn
Kim Jin
Hwang Chung-Ju
Abstract
Objective: The aim of this study was to compare recycled and unused orthodontic miniscrews to determine the feasibility of reuse. The comparisons included both miniscrews with machined surfaces (MS), and those with etched surfaces (ES).

Methods: Retrieved MS and ES were further divided into three subgroups according to the assigned recycling procedure: group A, air-water spray; group B, mechanical cleaning; and group C, mechanical and chemical cleaning. Unused screws were used as controls. Scanning electron microscopy, energy-dispersive X-ray spectrometry, insertion time and maximum insertion torque measurements in artificial bone, and biological responses in the form of periotest values (PTV), bone?implant contact ratio (BIC), and bone volume ratio (BV) were assessed.

Results: Morphological changes after recycling mainly occurred at the screw tip, and the cortical bone penetration success rate of recycled screws was lower than that of unused screws. Retrieved ES needed more thorough cleaning than retrieved MS to produce a surface composition similar to that of unused screws. There were no significant differences in PTV or BIC between recycled and unused screws, while the BV of the former was significantly lower than that of the latter (p < 0.05).

Conclusions: These resultsindicate that reuse of recycled orthodontic miniscrews may not be feasible from the biomechanical aspect.
KEYWORD
Retrieved orthodontic miniscrew, Recycling, Surface characteristics, Biomechanical characteristics
FullTexts / Linksout information
   
Listed journal information
SCI(E) ÇмúÁøÈïÀç´Ü(KCI) KoreaMed