Journal of Technologic Dentistry
2011 Volume.33 No. 1 p.55 ~ p.61
Mechanical Analysis of Macro-Hexagon Porous Dental Implant Using Selective Laser Melting Technique
Kim Bu-Sob

Choi Sung-Min
Abstract
Purpose: In this study, FEM(Finite Element Method) and bending strength test was conducted using normal implant and porous implant for the mechanical estimation of porous dental implant made by SLM method. Methods: Mechanical characteristics of PI(porous implant) and NI(normal implant) applied distributed loads(200N, 500N) were observed through FEM analysis. And each bending strength was gotten through bending test using MTS(Mechanical Test System, Instron 8871). Results: The result of FEM analysis was observed that stress difference between upper and surface of PI was 12 times, while NI was 2 times. The result of bending test was observed that bending strength of PI was lower than NI. we made a decision about this result that cross-sectional area of NI was larger than the PI. Conclusion: The stress shielding ability of porous implant was better than normal implant through result of FEM analysis. And bending strength of porous implant was lower than NI. We think that cause of this result was difference of cross-sectional area.
KEYWORD
SLM, dental implant, mechanical analysis, macro-hexagon porous dental implant