KMID : 0362120120340030213
|
|
Journal of Technologic Dentistry 2012 Volume.34 No. 3 p.213 ~ p.220
|
|
Three-dimensional evaluation on the repeatability and reproducibility of dental scanner-based digital models
|
|
Lee Gyeong-Tak
Kim Jae-Hong Kim Woong-Chul Kim Ji-Hwan
|
|
Abstract
|
|
|
Purpose: The aim of this study was to determine the repeatability and reproducibility of two dental scanners.
Methods: The master die and the stone replicas(Kavo, Germany) were digitized in touch-probe scanner(Incise, Renishaw, UK), white light scanner(Identica, Medit, Korea) to create 3-dimensional surface-models. The number of points in the point clouds from each reading were calculated and used as the CAD reference model(CRM). Discrepancies between the points in the 3-dimensional surface models and the corresponding CRM were measured by a matching-software(Power-Inspect R2, Delcam Plc, UK). The t-student test for one samples were used for statistical analysis.
Results: The reproducibility of both scanner was within , based on mean value. The mean value between measurements made directly on the touch probe scanner digital models and those made on the white light scanner digital models was , and was statistically significant(P<0.05).
Conclusion: With respect to adequate data acquisition, the reproducibility of dental scanner differs. Three-dimensional analysis can be applied to differential quality analysis of the manufacturing process as well as to evaluation of different analysis methods.
|
|
KEYWORD
|
|
CRM(CAD Reference Model), dental scanner, digital model, point cloud, repeatability, reproducibility
|
|