Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0362120180400010001
Journal of Technologic Dentistry
2018 Volume.40 No. 1 p.1 ~ p.8
Effect of drill radius setting on prosthesis machining
Kim Chong-Myeong

Kim So-Ri
Cho Mi-Hyang
Abstract
Purpose: The purpose of this study is to evaluate the trueness and fitness of machined prostheses according to drill radius setting in CAD software.

Methods: For this study, standard abutment were replicated in Type ¥³ stone. The stone abutment were scanned using a dental scanner. The CAD design software was designed using scanned abutment data. When designing, the drill radius was set to 0.3 mm and 0 mm, respectively, and saved. The saved design data was milled using a milling machine (n=13). The inner surface of the milled crown was scanned. The trueness and fitness were measured using the inner scan data of prostheses. Independent t-tests were performed to identify significant differences in each data.

Results: Trueness values of the data saved with 0.3 mm and 0 mm drill radius were 18.9 ¡¾2.3 §­ and 19.1 ¡¾0.9 §­, respectively. There was no statistically significant difference between the groups. Fitness values of the data saved with 0.3 mm and 0 mm drill radius were 65.5 ¡¾0.8 §­ and 33.8 ¡¾1.0 §­, respectively. There was a statistically significant difference between the groups (p<.05).

Conclusion : Setting the drill radius is important to produce clinically good fit prostheses.
KEYWORD
Dental CAD/CAM systems, Milling machine, Fitness, Three-dimensional evaluation, Trueness
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)