Abstract
Purpose: The purpose of this study was to evaluate and compare the accuracy of definitive casts that are fabricated from digital intraoral impression and conventional impression technique.

Methods: A master model(ANNA-4, Frasaco GmbH, Tettnang, Germany) with the prepared upper full arch tooth was used. Conventional impression and then stone model(n=10) were produced from this master model, and on the other hands, digital impressions were made with the CS-3600 intraoral scanner(n=10). Six linear measurements were recorded between landmarks, directly on each of the stone models on two occasions by a single examiner. Measurements were made with a digital caliper to the nearest 0.01mm from manual models and with the software(Delcam PowerSHAPE) from the virtual
models. The t-student test for paired samples and intraclass correlation coefficient(ICC) were used for statistical analysis. Results: The measurement of two methods showed good reliability. The ICC of the two models were 0.88~0.91(stone model) and 0.94~0.99(digital model). The mean differences to master model for stone model and digital model were 0.10~0.14mm, and 0.14~0.20mm, respectively.

Conclusion: The definitive casts obtained with digital intraoral technique model had significantly larger dimensions as compared to those of the stone model. However, the differences to the master model detected appear to provide enough accuracy and reliability for clinical application.