KMID : 0362320100350060453
Journal of Korean Academy of Operative Dentistry
2010 Volume.35 No. 6 p.453 ~ p.460
In vivo evaluation of accuracy and consistency of two electronic apex locators
Pi Chien-Yun

Kim Eui-Seong
Jung Il-Young
Lee Seung-Jong
Abstract
Objectives: To evaluate the accuracy and consistency of two different apex locators at both the Apex and 0.5 marks.

Materials and Methods : Twenty-six root canals was scheduled for extraction for periodontal or prosthodontic reasons. Thirteen canals were measured using Root ZX and the rest by i-ROOT. The root canal length was measured both the at 0.5 mark and the Apex mark. The file was then fixed to the toot, and the distance from the file tip to the major foramen of each canal was measured after removing the root dentin under the microscope so that the major foramen and the file tip were seen.

Results: 1. When the Apex mark was used, 100% of both the Root ZX and i-ROOT groups were within 0.5 mm of the major foramen. 2. When 0.5 mark was used, 100% of the Root ZX group and 77% of the i-ROOT group were within 0.5 mm of the major foramen. 3. In terms of standard deviation and quartile value, the Apex mark was more consistent than 0.5 mark in the Root ZX group, and 0.5 mark was more consistent in the i-ROOT group, but there was no statistically significant difference when compared with t-test. 4. The root canal length difference between the Apex mark and 0.5 mark was 0.22 mm and 0.46 mm in the Root ZX and i-ROOT groups, respectively.

Conclusions: In this study, the Apex mark was the more consistent mark. Therefore, it is recommended to subtract 0.5 mm, which is the average length between the apex and apical constriction, from the root canal length at the Apex mark to obtain the working length clinically.
KEYWORD
Accuracy, Consistency, Electronic apex locator, Microscope