KMID : 0362419960340030438
Journal of Korean Academy of Prosthodontics
1996 Volume.34 No. 3 p.438 ~ p.447
A study on failure strength of the implant prosthesis luter with resin cement between gold cylinder ad superstructure
Kim Seung-Beom

Bae Jeong-Sik
Abstract
This study was performed to determine the failure strength and pattern of implant prosthesis luted with resin cement between gold cylinder and superstructure, and to evaluate the bonding strength of resin cement. To evaluate failure strength and pattern, the groups were divided into 2. Group 2 : Casted gold cylinder Group 2 : Luted with resin cement between gold cylinder ans superstructure. To evaluate effects of the bonding strength of the implant prosthesis luted with resin cement according to storage condition, the groups were divided into 3 : Group A : Stored in waste at 37¡É for 24 hours. Group B : 1000 cycles thermocycled between 5¡É and 55¡É in water. Group C : Stored in water at 37¡É for 30 days.
@ES The results were as follows. @EN 1. Failure was found at gold screw in all specimens of group 1 & 2. 2. The bond strength in group 1 and 2 was 189.85 and 18.14Kgf. There was no statistically significant difference between group 1 and 2(P>0.05). 3. The bond strength of group A, B and C was 53.28, 45.86 and 39.29Kgf. There result suggest the advantage of an improved fit of superstructure to the abutment and a simple procedure. But there was a measurable decrease of the bond strength according to storage condition So, Further research is necessary to evaluate of the implant prosthesis luted with resin cement between gold cylinder and superstructure.
KEYWORD
FullTexts / Linksout information
  
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed