Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0362420060440060734
Journal of Korean Academy of Prosthodontics
2006 Volume.44 No. 6 p.734 ~ p.739
FATIGUE RESISTANCE OF DENTAL IMPLANTS TREATED WITH LASER METHOD
Park Eun-Young

Abstract
Statement of problem. Many surface processing methods of dental implant have been developed, the laser processing is one of them.

Purpose. This study was to investigate in vitro the fatigue resistance of implants treated with laser method(CSM implant, CSM Company, Daegu, Korea).

Material and methods. Fatigue tests of 23 laser treated CSM implant(CSM Company, Daegu, Korea) were conducted using Instron 8871 (Load cell, 5 kN, Instron Co., England), according to ISO 14801 (2003),

Results. From 300 N, each specimens were not fractured and withstood until more than 5,000,000 cycle.

Conclusion. Within the limits of this in vitro study, implants treated with laser method (CSM implant, CSM Company, Daegu, Korea) have had enough rigidity and fatigue resistance to use clinically with reliance.
KEYWORD
Fatigue resistance, Laser treated CSM implant, ISO 14801 (2003), Clinically, Reliance
FullTexts / Linksout information
   
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed