Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0362420080460010012
Journal of Korean Academy of Prosthodontics
2008 Volume.46 No. 1 p.12 ~ p.21
Comparative study in marginal adaptation of zirconia cores fabricated with 3 different CAD/CAM systems
Kwon Yong-Joong

Park Won-Hee
Lee Young-Soo
Abstract
Statement of problem: There have been many studies about marginal adaptation of Zirconia core fabricated with various CAD/CAM systems. But, the comparative study has not been conducted much.

Purpose: This study was to compare marginal adaptations of zirconia cores fabricated with 3 different CAD/CAM systems.

Material and methods: The in vitro marginal adaptations of zirconia cores fabricated with 3 different CAD/CAM systems(Adens system, Lava system, and Cerasys system) were evaluated and compared. In this study 30 zirconia cores were fabricated(10 zirconia cores per each system). The measurements of the marginal adaptatioin were carried out using microscope L-150A(Nikon, Japan). Measurements were recorded at 50 points. The measurements were analyzed using Kruskal-Wallis test and Wilcoxon rank sum test.

Results: Within the limits of this study, the results were as follows. 1. Mean marginal gap values
obtained were 284.5¡¾152.9§­ in Adens system, 99.0¡¾73.0§­ in Lava system, and 153.5¡¾85.7§­ in Cerasys system.
2. Zirconia core made by Lava system showed the smallest marginal gap(P<0.05).

Conclusion: In the limitation of this study, we concluded that Lava System was superior than other systems(Adens system and Cerasys system) in
the marginal adaptation.
KEYWORD
CAD/CAM system, Zirconia core, Marginal adaptation
FullTexts / Linksout information
   
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed