Abstract
Purpose : The purpose of this study was to evaluate the effect of amount of cantilever in intra-crown according to implant
fixture position on mechanical strength of internal conical joint type implant.

Materials and Methods : Internal conical joint type implant fixture, abutment screw, abutment was connected and gold alloy prostheses were fabricated and cemented on abutment. For fatigue fracture test, the specimens were loaded to the 350 N, 2,000,000 cycle on 3, 4, 5, and 6 mm off-center of gold alloy prostheses. The fracture pattern of implant component was observed.

Results : No fatigue fracture found on 3 and 4 mm group . But initial crack pattern found on 3 specimens of 4 mm group. Fatigue fracture found on all specimens of 5 mm group. But complete fracture was not observed. One specimen of 6 mm group fracture completely. Implant fixture fracture wax not observed.

Conclusion : The mechanical failure of implant prostheses increased with the loading area farther from center of implant fixture. To reduce mechanical problem of internal joint type implant, surgical and prosthetic consideration is needed.
KEYWORD
cantilever, fatigue test, internal conical joint type implant