Journal of the Korean Academy of Implant Dentistry
2004 Volume.23 No. 1 p.19 ~ p.31
A Finite Element Approach to Evaluate the Biomechanical Characteristics of Short Implants Placed Obliquely to the Occlusal Loads at the Vicinity of Maxillary Sinus
Yu Won-Jae

Lee Kyu-Bok
Abstract
Purpose: Investigation of the stresses around dental implants placed closed to the maxillary sinus which are subject to various loading conditions.

Materials and Methods: ITI solid implants of 6mm in length and 4.8mm in diameter placed at the molar region of maxilla were axisymmetrically modeled for a series of finite element analyses. Main consideration was given to the stresses at the cervical cortical bone induced by loads of 50N inclined to the implant axis as a result of inclined plantation of the implants. Based on the inclination angle between the load and the implant axis, seven different models were set for analyses. Resultant displacement and compressive stresses at the cervical bone were evaluated.

Results and Conslusions:
1. Higher stresses were observed at the cervical area of all the models.
2. The peak stresses at the cervical bone varied as a function of the inclination angle of the load. The bigger the angle, the higher the stress tended to occur.
3. Resultant displacement increased more rapidly in accordance with the inclination of the implant or the loading.
4. Stress of 3-4MPa was cautiously envisaged and assumed as the allowable level to the cyclic bone stress. Under this assumption, the inclination angle need to be controled to be within 5-13 degree for the present model. In the light of possible stress reduction followed by an adaption process at the cervical bone, however, a slightly bigger inclination might be accepted.
KEYWORD
ITI implant, Short implant, Inclined loading, Maxillary sinus, Cervical bone stress