KMID : 0362919880060010085
|
|
Journal of the Korean Society for Therapeutic Radiology and Oncology 1988 Volume.6 No. 1 p.85 ~ p.91
|
|
Dosimetric Characteristics of the KCCH Neutron Therapy Facility
|
|
Yoo Seong-Yul
Noh Sung-Woo Chung Hyun-Woo Cho Chul-Koo Koh Kyoung-Hwan Bak Joo-Shik Eenmaa Juri
|
|
Abstract
|
|
|
For the physical characterization of neutron beam, dosimetric measurements had been performed to obtain physical data of KCCH cyclotron-produced neutrons for clinical use. The results are presented and compared with the data of other institutions from the literatures. The central aixs percent depth dose, build-up curves and open and wedge isodose curve values are intermediate between that of a 4 and 6 MV X-ray. The build-up level of maximum dose was at 1.35 cm and entrance dose was approximately 40%. Flatness of the beam was 9% at Dmax and less than¡¾3% at the depth of 80% isodose line. Penumbra begond the 20% line is wider than corresponding photon beam. The output factors ranged 0.894 for 6 x 6 cm field to 1.187 for 30 x 30 cm field.
Gamma contamination of neutron beam was 4.9% at 2 cm depth in 10 x10 cm field.
|
|
KEYWORD
|
|
Neutron Beam, Dosimetry, cyclotron, Tissue equivalent chamber, Depth dose, Build-up, profile, Isodose curve, Output
|
|