KMID : 0362919890070010133
Journal of the Korean Society for Therapeutic Radiology and Oncology
1989 Volume.7 No. 1 p.133 ~ p.140
A Study of Three-dimension Tissue Equivalent Compensator for 6MV X-Rays
Kim Ok-Bae

Choi Tae-Jin
Suh Soo-Jhi
Abstract
Three-dimension paraffin compensator was designed to construct the tissue equivalent compensator for irregular body contours and obiliques beam incidence.
The ratio of compensator thickness to tissue deficit was depended on field size, depth and air gap because the scattered dose loss.
The ratio of compensator-tissue was optimized 0.79, 0.73, 0.61 and 0.50 in 6MV x-rays as function of field size 4¡¿4, 10¡¿10, 20¡¿20 and 30¡¿30 §² respectively in our study.
Using this tissue equivalent compensator, it can be got 2% difference of dose at same mid-plane in phantom study.
KEYWORD
3-D compensator, tissue deficit, TL Dosimetry
FullTexts / Linksout information
Listed journal information