Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0362919900080010125
Journal of the Korean Society for Therapeutic Radiology and Oncology
1990 Volume.8 No. 1 p.125 ~ p.131
Dose Attenuation in the Mid-Cranial Fossa with 6 MV hoton Beam Irradiations
Park Jeong-Ho

Choi Tae-Jin
Kim Ok-Bae
Abstract
In X-ray irradiation, dose distribution depends on multiple parameters, one of them being tissue inhomogeneity to change the dose significantly. Cpnsoderable dose attenuation through the mid-cranial fossa is expected because of various bony structures in it.
Dose distribution around the mid-cranial fossa, following irradiation with 6 MV photon beam, was measured with LiF TLD micro-rod, and compared with the expected dose in the same sites.
In our calculation with Cf(correction factor), the expected dose attenuation revealed about 3.74% per 1 cm thickness of bone tissue. And the differences between the expected dose with correction for bone tissue and the measured dose by TLD was small, agreeing within an average variation of ¡¾0.21%.
KEYWORD
Bone correction factor, TLD, Mid-cranial fossa
FullTexts / Linksout information
Listed journal information