KMID : 0362919910090020337
|
|
Journal of the Korean Society for Therapeutic Radiology and Oncology 1991 Volume.9 No. 2 p.337 ~ p.342
|
|
The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and C¥ë / CE Method
|
|
Yi Byong-Yong
Choi Eun-Kyung Chang Hye-Sook
|
|
Abstract
|
|
|
The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and C¥ë / CE (ICRU-21 and SCRAD protocol) method is studied. The therapetutic range of radiation (photon 4MV-15MV and electron 6MeV-20MeV)and three kinds of the chambers were used in the water phantom. The Results from TG-21 and KAPM Protocol did not show much differences (less than 1%) throughout the whole energy range; ND from KAPM protocol and Ngas from TG-21 showed 0.2% deviation mainly from W/e difference between two protocols. But the results from KAPM protocol (1990) and those from C¥ë / CE Method showed –1.9¡¾0.6%(KAPM protocol is higher) deviation for photom beam and £«3.3¡¾1% (KAPM protocol is lower) deviation for electron beams.
|
|
KEYWORD
|
|
Dosimetric Protocol, KAPM Dosimetric Protocol, TG-21
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|