Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0362919920100020147
Journal of the Korean Society for Therapeutic Radiology and Oncology
1992 Volume.10 No. 2 p.147 ~ p.153
Dose Effect of Tissue Compensator for 6 MV X-Ray
Lee Ho-Jun

Choi Tae-Jin
Kim Ok-Bae
Abstract
It is ideal thing to compensate tissue deficit without skin contamination in curvatured irradiation field of high energy Photon beam. The 3-dimensional compensating technique utilizing tissue equivalent materials to ensure an adequate dose distribution and skin sparing effect was described. This compensator was made of paraffin(70%) and stearin wax (30%) compound. The parameters for evaluation of the effect on skin dose in application of compensator were considered in the size of the field, the thickness of the compensator and the source-to-axis distance.
The results are as follows; the skin doses were not changed even though application of the compensator, but depended on the field size and the source-to-axis distance, and the skin doses were only slightly changed within 1% relative errors as increasing the thickness of the compensator in these experiments.
KEYWORD
3-dimensional compensator, Tissue equivalent material, Skin dose