Journal of the Korean Society for Therapeutic Radiology and Oncology
1993 Volume.11 No. 2 p.449 ~ p.454
The Enhancement of Skin Sparing by Tray Materials for High Energy Photon Beam
Chu Sung-Sil

Lee Chang-Geol
Kim Gwi-Eon
Abstract
The skin sparing effect associated with high energy x-ray or gamma ray beams may be reduce or lost under certain conditions of treatment. Current trends in using large fields. Shield carrying trays, compensating filters, and isocentric methods of treatment have posed problems of increased skin dose which sometimes become a limiting factor in giving adquate tumor doses. We used the shallow ion chamber to measure the phantom surface dose and the physical treatment variables for Co-60 gamma ray, 4MV and 10 MV x-ray beam. The dependence of percent surface dose on field sizes, atomic number of the shielding tray materials and its distance from the surface for 4, 10MV x-rays and Co-60 gamma ray is qualitatively similar. The use of 2 mm thick tin filter is recommended for situations where a low atomic number tray is introduced into the beam at distances less than 15 cm from the surface and with the large field sized for 4 MV x-ray beam. In case of Co-50 gamma ray, the lead glass tray is suitable for enhancement of skin sparing. Also, the filter distance should be as large as possible to achieve substantial skin sparing.
KEYWORD
Skin sparing effect, Tin filter, Lead glass, Skin dose