Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0362919940120020219
Journal of the Korean Society for Therapeutic Radiology and Oncology
1994 Volume.12 No. 2 p.219 ~ p.223
Total Body Irradiation Technique: Basic Data Measurements and In Vivo Dosimetry
Choi Dong-Rak

Choi Ihl-Bohng
Kang Ki-Mun
Shinn Kyung-Sub
Kim Choon-Choo
Abstract
This paper describes the basic date measurements for total body irradiation with 6 Mv photon beam including compensators designs. The technique uses bilateral opposing field with tissue compensators for the head, neck, lungs, and legs from the hip to toes.
In vivo dosimetry was carried out for determining absorbed dose at various regions in 7 patients using diode detectors (MULTIDOSE, Model 9310, MULTIDATA Co., USA). As a results, the dose uniformity of¡¾3.5%(generally, within¡¾10%) can be achieved with our total body irradiation technique.
KEYWORD
Total body irradiation, Tissue compensator, In vivo dosimetry
FullTexts / Linksout information