KMID : 0363020120420060243
Journal of Korean Academy of Periodontology
2012 Volume.42 No. 6 p.243 ~ p.247
A comparison of magnetostrictive and piezoelectric ultrasonic scaling devices: an in vitro study
Yousefimanesh Hojatollah

Robati Maryam
Kadkhodazadeh Mahdi
Molla Reza
Abstract
Purpose: The effects of magnetostrictive and piezoelectric devices on tooth surfaces seem to differ with regard to the root surface roughness they produce. This study aimed to compare the results of scaling using magnetostrictive and piezoelectric devices on extracted teeth.

Methods: Forty-four human extracted teeth were assigned to four study groups (n=11). In two groups (C100 and C200), the teeth were scaled using a magnetostrictive device and two different lateral forces: 100 g and 200 g, respectively. In the other two groups (P100 and P200), the teeth were scaled with a piezoelectric device with 100 g and 200 g of lateral force, respectively. he teeth were scaled and the data on the duration of scaling and the amount of surface were collected and analyzed using the t-test.

Results: The mean time needed for instrumentation for the piezoelectric and magnetostrictive devices was 50:54 and 41:10, respectively, but their difference was not statistically significant (P=0.171). For root surface roughness, we only found a statistically significantly poorer result for the C200 group in comparison to the P200 group (P=0.033).

Conclusions: This study revealed that applying a piezoelectric scaler with 200 g of lateral force leaves smoother surfaces than a magnetostrictive device with the same lateral force.
KEYWORD
Dental calculus, Dental scaling, Tooth extraction