KMID : 0364020170500030224
|
|
Korean Journal of Thoracic and Cardiovascular Surgery 2017 Volume.50 No. 3 p.224 ~ p.227
|
|
Extravascular Migration of a Fractured Inferior Vena Cava Filter Strut
|
|
Lim Jung-Hyeon
Lee Weon-Yong Ra Yong-Joon Jeong Jae-Han Park Bong-Suk Ko Ho-Hyun
|
|
Abstract
|
|
|
A 20-year-old man presented with a femur fracture and epidural hemorrhage (EDH) following a fall. One month after fracture surgery, swelling developed in both legs, and he was diagnosed as having a deep-vein thrombosis and pulmonary embolism. A retrievable inferior vena cava filter (IVCF) was inserted, because EDH is a contraindication to anticoagulants. Four months later, he complained of abdominal pain, and a computed tomography scan showed a fractured IVCF strut. After percutaneous removal failed 3 times, the IVCF was surgically removed by orthopedists using a portable image intensifier without cardiopulmonary bypass.
|
|
KEYWORD
|
|
Vena cava filters, Migration, Fracture
|
|