Journal of Radiation Protection and Research
2005 Volume.30 No. 4 p.231 ~ p.236
Measurement of ECF for CaSO_4: Dy Thermoluminescent Dosimeters
Lim Kil-Sung

Kim Jang-Ryul
Abstract
Dosimeters are manufactured from same process in the manufacturer but the deviation of TL raw counts exists among the dosimeters. TL raw counts are also gradually degrade due to multiple readings and physical abuse. ECF (Element Correction Factor) correct the degradation and deviation of TL raw counts to the average TL raw counts of reference dosimeters. Procedures for producing ECF of thermoluminescent dosimeters were described In detail. ECFs of 319 reference, control and field dosimeters were measured three times and average of three ECF values was calculated. Also, % CV(Coefficient of Variation) of three ECF values was calculated to verify ECF. ECF & % CV distributions for the field and control dosimeters are presented. TL raw counts of field dosimeters, being used about 6 times for the past 3 years, were almost unchanged, but those of control dosimeters being used more frequently, were degraded about 4.7 %.
KEYWORD
element correction factor(ECF), themoluminescent dosimeter(TLD), calibration