Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0364920090340020043
Journal of Radiation Protection and Research
2009 Volume.34 No. 2 p.43 ~ p.48
POLARITY AND ION RECOMBINATION CORRECTION FACTORS OF A THIMBLE TYPE IONIZATION CHAMBER WITH DEPTH IN WATER IN THE MEGAVOLTAGE BEAMS
Kim Seong-Hoon

Huh Hyun-Do
Choi Sang-Hyoun
Min Chul-Hee
Shin Dong-Oh
Choi Jin-Ho
Abstract
When the PDD (percentage depth dose) in the megavoltage beams is measured in the water phantom, the polarity and ion recombination effects of ionization chambers with depth in water are not usually taken into consideration. We try to investigate if those variations with depth should be taken into consideration or could be ignored for the thimble type semiflex ionization chamber (PTW , SN 1551). According to the recommendation of IAEA TRS-398, the 4 representative depths of , , and were used for the electron beams. For the photon beams, the 4 depths were arbitrarily chosen for the photon beams, which were , , and . For the high energy photon beam both polarity and ion recombination factors of the chamber with depth in water gives the good agreements within the maximum 0.2%, while the with depth came within the maximum 0.4% and the within the maximum 0.6% in every electron beam used. This study shows that PDI (percentage depth ionization) could be a good approximation to PDD for the chamber used.
KEYWORD
Polarity Effect, Ion Recombination Effects, Ion Chamber, PDD, PDI
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)