Journal of Radiation Protection and Research 2010 Volume.35 No. 2 p.81 ~ p.84
|
|
Evaluation of a Fabricated Charge Sensitive Amplifier for a Semiconductor Radiation Detector
|
|
Kim Han-Soo
Ha Jang-Ho Park Se-Hwan Lee Jae-Hyung Lee Cheol-Ho
|
|
Abstract
|
|
|
A CSA(Charge Sensitive Amplifier) was designed and fabricated for application in a radiation detection system based on a semiconductor detector such as Si, SiC, CdZnTe and etc.. A fabricated hybrid.type CSA was evaluated by comparison with a commercially available CSA. A comparison was performed by using calculation of ENC (Equivalent Noise Charge) and by using energy resolutions of fabricated radiation detectors based on Si. In energy resolution comparison, a fabricated CSA showed almost the same performance compared with a commercial one. In this study, feasibility of a fabricated CSA was discussed.
|
|
KEYWORD
|
|
CSA (Charge Sensitive Amplifier), ENC (Equivalent Noise Charge), Energy resolution, Semiconductor radiation detector
|
|
FullTexts / Linksout information
|
|
|
|