Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0364920100350020081
Journal of Radiation Protection and Research
2010 Volume.35 No. 2 p.81 ~ p.84
Evaluation of a Fabricated Charge Sensitive Amplifier for a Semiconductor Radiation Detector
Kim Han-Soo

Ha Jang-Ho
Park Se-Hwan
Lee Jae-Hyung
Lee Cheol-Ho
Abstract
A CSA(Charge Sensitive Amplifier) was designed and fabricated for application in a radiation detection system based on a semiconductor detector such as Si, SiC, CdZnTe and etc.. A fabricated hybrid.type CSA was evaluated by comparison with a commercially available CSA. A comparison was performed by using calculation of ENC (Equivalent Noise Charge) and by using energy resolutions of fabricated radiation detectors based on Si. In energy resolution comparison, a fabricated CSA showed almost the same performance compared with a commercial one. In this study, feasibility of a fabricated CSA was discussed.
KEYWORD
CSA (Charge Sensitive Amplifier), ENC (Equivalent Noise Charge), Energy resolution, Semiconductor radiation detector
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)