Manufacture of Immobilization Device (Vac-lok) and the Usefulness of Evaluation for Reproducibility of Patients in the Radiotherapy
|
|
Lee Jae-Seung
Kim Gyoung-Hee Jeong Kyu-Hwan Kweon Dae-Cheol Goo Eun-Hoe Park Cheol-Woo Im In-Chul Dong Kyung-Rae Chung Woon-Kwan
|
|
Abstract
|
|
|
We manufactured the Vac-lok that can be applied to patient of special body shape and need to special set-up position and evaluated the usefulness in the radiation therapy. The manufacture Vac-lok, It was used EVA resin and biobeadform of a diameter 1.5 mm. carried out the test of functionality, structural and analyzed the relative reproducibility of phantoms and patients. During the total period of radiation therapy, Vacuum pressure bring variety to a very small amount in the test of functionality of the manufacture Vac-lok. But It was a negligible quantity. The manufacture Vac-lok improved the relative reproducibility of phantoms than the existing Vac-lok and tolerance has a confidence less than 4% error. Also, relative reproducibility of patient increased error than phantom in the antero-posterior and lateral plan. However, the maximum set-up error was less than ¡¾ 2.3 mm. In conclusion, If tolerance set-up error of radiotherapy is less than ¡¾ 3 mm, the manufacture Vac-lok was enough possible to use and improvement of reproducibility, considering supply with the Vac-lok made to measure of special patient that produced at a low price and without delay.
|
|
KEYWORD
|
|
Vac-lok, radiation therapy, EVA resin, biobeadform, reproducibility
|
|