Development of the Graphite-Moderated Neutron Calibration Fields Using 241Am-Be Sources in JAEA-FRS
Nishino Sho

Tanimura Yoshihiko
Ebata Yoshiaki
Yoshizawa Michio
Abstract
Background: The moderated neutron calibration fields using 241Am-Be sources and a graphite moderator have been constructed at the Facility of Radiation Standard (FRS) in the Japan Atomic Energy Agency (JAEA).

Materials and Methods: The neutron spectra of the fields were evaluated by the Monte-Carlo calculations and measurements using the Bonner Multi-sphere Spectrometer.

Results and Discussion: The fields have continuous neutron spectra from several MeV to thermal neutron energy, with fluence-averaged energies of 0.84 MeV and 0.60 MeV. Reference values of fluence rates and ambient/personal dose equivalent rates were determined from neutron spectra by measurements.

Conclusion: Currently, the fields are available for calibration or performance test of neutron measuring instruments.
KEYWORD
Moderated neutron calibration field, 241Am-Be source, FRS, Neutron dosimetry