KMID : 0371320120830030183
|
|
Journal of the Korean Surgical Society 2012 Volume.83 No. 3 p.183 ~ p.186
|
|
Stent fractures after superficial femoral artery stenting
|
|
Park Jae-Young
Jeon Yong-Sun Cho Soon-Koo Jin Chan-Ik Kim Kyung-Rae Shin Woo-Young Kim Jang-Young Hong Kee-Chun
|
|
Abstract
|
|
|
Stent fracture is one of the major factors compromising implanted stent patency due to its consequences including in-stent restenosis, thrombosis, perforation, and migration. Stent fracture can occur from stress (extrinsic or intrinsic) and biomechanical forces at different implantation sites. We report on 2 cases of stent fractures and pertinent literature. One patient, a 75-year-old male, presented with recurrence of claudication 14 months after superficial femoral artery stenting; a femoral artery occlusion with stent fracture was found, and he underwent femoropopliteal bypass. The other patient, a 72-year-old male presented with recurrence of claudication; a stent fracture was found without femoral artery occlusion, and he was treated with additional femoral artery stenting to secure the fracture site.
|
|
KEYWORD
|
|
Stents, Vascular patency, Femoral artery, Early intervention
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|