Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0371619960120010065
Journal of Wonkwang Medical Science
1996 Volume.12 No. 1 p.65 ~ p.74
Development of Wonkwang University Type LINAC-Based Stereotactic Radiosurgery System



Abstract
The purpose of this study is to develop LINAC based stereotactic radiosurgery system using ML-6M type linear accelerator which had been installed in Wonkwang University Hospital. Mechanical alignment, physical beam distribution and treatment
simulation
were evaluated. All the mechanical and physical data were acceptable and safe for clinical application. In the treatment simulation. Good agreement between the center of the radiation distribution and the localized targed point was achieved. In
conclusion. This type of radiosurgery system can be applied to clinical setting with acceptable accuracy and safety.
KEYWORD
FullTexts / Linksout information
Listed journal information