Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919820030010043
Journal of Biomedical Engineering Research
1982 Volume.3 No. 1 p.43 ~ p.48


Abstract
Ventricular assist device(VAD) has been clinically applied as a temporary circulatory sup- porting system in the patients with severe heart failure, but small sized VAD for infant is not available. The purpose of this paper is to introduce small sized VAD and presents the result of in vitro test. Sejong VAD is diaphragmatic type of pneumatic pump and stroke volume is 11cc. Cardiac outputs of the Sejong VAD were measured by overflow tank under variable conditions of driving parameters. The cardiac output was 1.3 1/min at the heart rate of 120 per minute, left atrial pressure of 15cmH2O, percent systole of 43%, driving pressure of 240 mmHg, vacuum pressure of -40 mmHg, and mean aortic pressure of 70 mmHg. No mechanical problem was developed during the continuous in vitro test for 3 months.
KEYWORD
FullTexts / Linksout information
Listed journal information