Journal of Biomedical Engineering Research
1982 Volume.3 No. 2 p.95 ~ p.100
Glucose Analysis Using Free and Immobilized Glucose Oxidase Electrode



Abstract
Glucose oxidase from A. niger was entrapped in polyacrylamide gel which was used in the enzyme electrode for glucose analysis. The electrode was assembled by placing the gel between the membranes on the surface of a Clark type electrode. In order to make it possible to analyze the experimental results later, the stagnation flow was adopted wheree the governing fluid mechanics were well known. The current increased with the increase concentration in the bulk below a certain level of glucose concentration beyond which no more current increase was observed. This is probably due to the diffusion limitation of oxygen from the bulk solution. Also the current increased witll the enzyme loading in the gel, but the linearity between the current and the glucose concentration was rather limited to a narrow range. Flow rate was found to be very important, which means that film diffusion is very important under the flow rate of 5cm/sec. As a conclusion, enzyme loading, gel layer thickness, stirring speed and bulk concentration of glucose were found to be most improtant parameters in yielding a linar current reponse with respect to the bulk glucose concentration.