Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919870080020171
Journal of Biomedical Engineering Research
1987 Volume.8 No. 2 p.171 ~ p.176
Image Restoration in Dual Energy Digital Radiography using Wiener Filtering Method.
¹Îº´±¸
¹Ú±¤¼®
Abstract
Wiener filtering method was applied to the dual energy imaging procedure in digital radiography(D.R.). A linear scanning photodiode arrays with 1024 elements(0.6mm H 1.3mm pixel size) were used to obtain chest images in 0.7 sec. For high energy image acquisition, X-ray tube was set at 140KVp, 100mA with a rare-earth phosphor screen. Low energy image was obtained with X-ray tube setting at 70KVp, 150mA. These measured dual energy images are represented in the vector matrix notation as a linear discrete model including the additive random noise. Then, the object images are restored in the minimum mean square error sense using Wiener filtering method in the transformed domain. These restored high and low energy images are used for computation of the basis image decomposition. Then the basis images are linearly combined to produce bone or tissue selective images. Using this process, we could improve the signal to noise ratio characteristics in the material selective images.
KEYWORD
FullTexts / Linksout information
Listed journal information