Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919890100030309
Journal of Biomedical Engineering Research
1989 Volume.10 No. 3 p.309 ~ p.315
Development of the Patient Monitor Using Microprocessor


Yoo Sun-Kook

Abstract
In this paper, the patient monitor consisting of amplifier, scan converter, A/D converter, CRT amplifier, and micro-controller part was developed. This patient monitor measures the patient¡¯¡¯¡¯¡¯s 4 states in the hospital such as electro-cardiography, respiration, blood pressure, and temperature. The control and processing methods based on micro-processor employ the flexibility, extensibility and economy over other conventional system. The followings are incorporated in this system. First, record the heart rate trends for 1 and 4 hours respectively. Second, measures the respiration by impedance pneumography. Third, measures the blood pressure with auto-zero balance. Fourth, linesrize the temperatures by bridge method.
KEYWORD
FullTexts / Linksout information
Listed journal information