Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919900110010031
Journal of Biomedical Engineering Research
1990 Volume.11 No. 1 p.31 ~ p.37
Correction of Ocular Fundus Fluorescein Angiogram in Nonlinear Distortion


Abstract
Image correction of the fluorescein angiogram with the low contrast and geometrically distorted is studied. The procedure is divided into two steps : A template matching step and a geometrical transformation step. Based on the sequential similiarity detection algorithm, more Improved matching results are obtained with edge enhanced image. For the reference points and the corresponding matchimg points, the geometrical transformationss by the use of the average value of the X, Y translational shifts and a pair of these points, respectively are described. Then the corrected images are obtained using bilinear interpolation. It is proven that the usefullness of this study for the image correction of the ocular fundus fluoresceln angiogram with low contrast and geometrically distorted mainly due to the eye movements
KEYWORD
FullTexts / Linksout information
Listed journal information