Journal of Biomedical Engineering Research
1990 Volume.11 No. 1 p.113 ~ p.119
Resolution Enhancement of Ultrasonic B-scan Images by Modified Wiener Filter
Á¤ÁØ¿µ
Áø¿µ¹Î/À̳뼺/¿ì±¤¹æ/À¯Çü½Ä
Abstract
In this paper, the deconvolution method utilizing a modified Wiener filter is applied for the enhancement of lateral resolution of ultrasonic B-scan Images. For this purpose, a phantom composed of wires which are 0.6mm of diameter and apart in the range between 3 to 9mm is constructed. The modified Wiener filter with optimal parameter is applied to the phantom for the analysis of ultrasonic image. The results obtained are as follows¡¯¡¯¡¯¡¯When all parameters of the modified Wiener filter are optimal, the resolution of B-scan images is enhanced by 50 percent : Othenrise, the images are blurred, spilt at peak points, or noises are strengthened severely. When the point-spread function representing the characteristic function of the system is determined, the selection ranges of op- timum parameters may be narrowed. It is expected that the proposed method may be able to apply to clinic situations for more accurate image analysis by means of reducing the loss of important information.
KEYWORD