Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919900110010147
Journal of Biomedical Engineering Research
1990 Volume.11 No. 1 p.147 ~ p.156
A Study on the Implementation Simulation and System for 2-D Doppler System Using Second-order Sampling




Abstract
A two-dimensional pulsed doppler system for ultrasonic blood velocity doppler signals is studied and implemented. The second-order sampling method and serial data processing procedures are utillized in the sys- tem, which eliminates the untuning problems at phase channels in the quadrature detection method as well as in the channels of parallel data processing. rho digital signal processor used in this system allows a hardware savings and flexible design options. The efficiency of the various mean frequency estimators in the second-order sampling system is examined by computer simulation as a function of the intersequence sample delay time. The temporal delay for the quadrature component is changed from 1/(4fo) to 3/(4fo) and 5/(4fo) where to is the center frequency of the transducer, It is found that autocorrelator is the optimum frequency estimator for the second-order sampling: with !he intersequence sample delay of 1/(4fo) to 3/(4fo) and 5/(4fo). The qualitative variation and information proportional to blood velocity in the vessel system are obtained in the VIVO experiments.
KEYWORD
FullTexts / Linksout information
Listed journal information