Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919910120010035
Journal of Biomedical Engineering Research
1991 Volume.12 No. 1 p.35 ~ p.42
The Analysis of the Information Capacity of Digital Radiography System




Abstract
In this paper, the information capacity, the intergrate performance measure of digital radiography system has been quantitatively analyzed. The effect of various factors affecting the information capacity of the digital radiography system in multistage detection processes has been considered and analyzed In detail. The results show that the more information capacity can be obtained with the smaller detector cell area, despite of the reduced signal to noise ratio due to the reduced number of X ray photons per detector cell. Considering the limit of human visual acuity, however. the sufficient resolution will be obtained when the detector cell size in 0.2x0.2mm2 with 8 bit quantizaion. And also the results indicates that the information capacity may be severely reduced by the mixture of electric noise in final read-out stage.
KEYWORD
FullTexts / Linksout information
Listed journal information