KMID : 0372919910120030185
|
|
Journal of Biomedical Engineering Research 1991 Volume.12 No. 3 p.185 ~ p.190
|
|
Mesurement of Evoked Otoacoustic Emission Latency Using Linear prediction Coding Spectrum
|
|
|
|
Abstract
|
|
|
An automatized latency calculation method of an e-OAE(evoked otacoustic emission) is proposed. The e-OAE signal measured from a normal adult is averaged 1000 times to remove noises. This averaged signal is converted to digital signal and that is processed by IBM-AT computer for latency calculation. we separate the stimulated and the emitted signal on the time domain by a modified LPC (linear prediction coding) spectrum, and the latency is calculated by cross-correlation method. By proposed latency calculation method the latency is 7.9[ms] for normal adult. The performance of the proposed method is also compared with that of the auto-correlation and cross performance of the proposed method is also compared with that of the auto-correlation and cross-correlation method. The result show that the proposed method has same precision with the conventional methods and can automatically calculate latency without subjective observation.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|