Mesurement of Evoked Otoacoustic Emission Latency Using Linear prediction Coding Spectrum



Abstract
An automatized latency calculation method of an e-OAE(evoked otacoustic emission) is proposed. The e-OAE signal measured from a normal adult is averaged 1000 times to remove noises. This averaged signal is converted to digital signal and that is processed by IBM-AT computer for latency calculation. we separate the stimulated and the emitted signal on the time domain by a modified LPC (linear prediction coding) spectrum, and the latency is calculated by cross-correlation method. By proposed latency calculation method the latency is 7.9[ms] for normal adult. The performance of the proposed method is also compared with that of the auto-correlation and cross performance of the proposed method is also compared with that of the auto-correlation and cross-correlation method. The result show that the proposed method has same precision with the conventional methods and can automatically calculate latency without subjective observation.