A Design of New Surface Gradient Coil and Its Application to MR Computerized Tomography
Yi JH
Cho ZH
Abstract
A new three-channel surface gradient coil obtained by using numerical optimization and its application to MR computerized nomography are presented. The new surface gradient coil pro aided linear field gradient region more than twice wider compared with the t¡¯¡¯¡¯¡¯irst surface gradi encl coil, removed torque and field offset, and reduced coupling between the surface gradient coil and combined surface rf coil. We realized the new surface gradient coil set with 30 x 60cm2 size, which generated more tharl 4G/cm with 100 amperes over a 4x4x4cm3 region with good linearity. The optimal geometries of the three-channel surface gradient coil and volun teer¡¯¡¯¡¯¡¯s high-resolution in wiuo spinal cord Images obtained by using the optimized surface gradi ent coil set are presented.