Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919950160040447
Journal of Biomedical Engineering Research
1995 Volume.16 No. 4 p.447 ~ p.456
Development of Transcutaneous Energy Transmission System for Medical Instrument




Abstract
As a part of electro-mechanical totally implantable artificial heart (TIAH) program, a transcut- aneous energy transmission system has been developed. By mutual magnetic induction between the first coil on the skin and the subcutaneously implanted second coil, the system transfers elctrical power through the skin. This research aimed at minimizing the size of the implanted part as well as maximizing the transfer efficiency. Using class I amplifier, we achieved above 75% power transfer efficiency at average 40W power transfer level which is required for normal TIAH operation. In vivo performance of the developed system and bio-compatibility of the material used in Implanted parts were evaluated through animal experiments.
KEYWORD
FullTexts / Linksout information
Listed journal information